S-623 Precise Conductivity/DO Meter

For high-precision measurements of the conductivity/DO in laboratory. Widely used in the fields of biotechnology, aquaculture, power plants and environmental protection.

Main Features:

●6.5 inch LCD screen which is intuitive and can show more parameters. Humanized design of operating interface makes operations simpler.

●Omron long lasting and light touch panel can be used more than 100,000 times.

● Meet GLP norms, with auto calibration, auto temperature compensation, data storage, USB data export, clock, wireless printing, function setup, smart self-diagnosis and other intelligent functions.

●Built-in big memory can store 500 sets of test data which can be saved and transferred to USB flash drive and opened with Excel, very easy and convenient for edit.

●Auto recognition of eight kinds of standard conductivity solution and there are two series of standard solution for option: Europe & USA series and China series.

●Adopt advanced conductivity measurement technology, one point calibration of conductivity electrode with constant K=1 can meet the measuring accuracy requirements for test 0.5μ S/cm ~ 200mS/cm.

● Able to switch between conductivity, TDS, salinity and resistivity. Multinomial calculation for TDS and salinity to ensure the conversion precision of the full scale.

●DO mode has functions of ATC, auto salinity compensation and auto pressure compensation.

●New structure polarographic DO electrode can measure temperature, salinity and dissolved oxygen simultaneously, equipped with special DO electrode calibration sleeve, only need 3~5 minutes for polarization.

●Combined diaphragm cap makes it extremely convenient to replace the sensitive membrane just by screwing the new cap to the electrode.

●With standard Bluetooth block and support wireless Bluetooth printing to make operations much easier for users. Optional data transmission to PC and cell phone via wireless Bluetooth.

● IP54 dustproof and waterproof housing

Conductivity Measuring range

 

 

 

 

 

 

 

Resolution

Accuracy

Temp compensation

Electrode constant

Reference temperature

Conductivity:

(0.00~20.00) μS/cm  (20.0~200.0) μS/cm

(200~2000) μS/cm    (2.00~20.00) mS/cm

(20.0~200.0) mS/cm

(200~2000) mS/cm (K=10)

Resistivity: (0 ~ 100) MΩ·cm

TDS: (0 ~ 100) g/L

Salinity: (0 ~ 100) ppt

0.01/0.1/1μS/cm   0.01/0.1/1mS/cm

Electrode: ±0.5% FS, Instrument: ±0.80%FS

(0 ~ 50) ℃ (automatic)

0.1/0.5/1/5/10/50/100 cm-1

25℃, 20℃,18℃

Dissolved Oxygen Measuring range

Resolution

Accuracy

Response time

Residual current

Temperature compensation

Salinity compensation

Pressure compensation

Automatic calibration

Electrode type

(0 ~ 20.00) mg/L(ppm),  (0 ~ 200.0)%

0.1/0.01 mg/L(ppm),  1/0.1 %

Electrode: ±0.10 mg/L, Instrument: ±0.40 mg/L

≤30s (25℃, 90% response)

≤0.1 mg/L

(0 ~ 45)℃ (automatic)

(0 ~ 45) ppt (automatic)

(80 ~ 105) kPa (automatic)

Air saturated by water, water saturated by air

Polarographic

Temperature Measuring range

Resolution

Accuracy

0℃~ 100℃

0.1℃

5~ 60℃: ±0.4℃   Others: ±0.8℃

Other Parameters Data storage

Power

Communications

Dimensions & Weight

500 groups

DC12V/1A

RS232, USB2.0, wireless Bluetooth

270×170×70mm/600g

Working Conditions

 

Ambient temperature

Ambient humidity

IP grade

10 ~ 30℃

≤ 85%

IP54