P-512 Dissolved Oxygen Meter

Suitable for aquaculture, water source monitoring, sewage treatment and environmental protection monitoring, especially for outdoor purposes.

●Large LCD screen with controllable white backlight.

●Humanized designs like rack and skid proof shell make it more convenient for users.

● DO mode has functions of ATC, auto salinity compensation and auto pressure compensation.

● New structure polarographic DO electrode can measure temperature, salinity and dissolved oxygen simultaneously, equipped with special DO electrode calibration sleeve, only need 3~5 minutes for polarization.

●Built-in microprocessor chip with functions of auto calibration, auto temperature compensation, data storage, function setup, self-diagnosis, auto power off, low voltage display.

●Combined diaphragm cap makes it extremely convenient to replace the sensitive membrane just by screwing the new cap to the electrode.

●Grade IP57 and equipped with portable carrying case.

Specifications

Dissolved Oxygen Measuring range (0 ~ 20.00) mg/L(ppm), (0 ~ 200.0)%
Resolution 0.1/0.01 mg/L(ppm), 1/0.1 %
Accuracy Electrode: ±0.10 mg/L, Instrument: ±0.40 mg/L
Response time ≤30s (25℃, 90% response)
Residual current ≤0.1 mg/L
Temperature compensation (0 ~ 45)℃ (automatic))
Salinity compensation (0 ~ 45) ppt (automatic)
Pressure compensation (66 ~ 200) kPa (manual)
Automatic calibration Air saturated by water, water saturated by air
Electrode type Polarographic
Other Technical Parameters Data storage 100 groups
Storage content Series number, measuring value measuring unit and temperature,
Power Two AA batteries(1.5V×2)
Dimensions & Weight 65×120×31mm/180g (instrument)

255×210×50mm/790g (with carrying case)

Working Conditions Ambient temperature (5 ~ 35)℃ (grade 0.01)
Ambient humidity ≤85%
IP grade IP57
Scroll to Top