Specification
Chlorine Sensor Probe
Type: Membrane covered potentiostatic chrono amperometric three-electrode system
Measurand: Total Chlorine or Free Chlorine
Range: 0.01-2, 0.01-5, 0.01-10mg/l, 0.01-20mg/l, 0-200mg/l (ppm)
Resolution: 0.01mg/l (ppm)
Reproducibility: ±5 %
Stability: –2 % per month (without calibration)
Working electrode: Gold
Counter electrode: Stainless Steel
Reference electrode: Silver / silver halide
Membrane material: Micro-porous hydrophilic membrane
Flow rate: Approximately 0.5l/min (min 0.2l/min)
Temperature range: >3 to <50 oC
Temperature compensation: Automatically by an integrated thermistor
pHrange: pH 4 up to pH 9
Response time: T90: approx.120 seconds
Calibration: Manual using DPD
Housing material: PVC, silicone, polycarbonate stainless steel
Maintenance intervals:
Membrane: 12-18 months
Electrolyte: 3-6 months
Interferences: High levels of other oxidants such as Ozone and Chlorine Dioxide